To compare corneal pachymetry values measured by three different optical devices: Orbscan II, Pentacam HR and Sirius in healthy eyes. METHODS: The central corneal thickness (CCT) of 102 eyes of 102 healthy subjects (mean age of 33.09±8.72 years and mean refractive defect -4.11±4.74 D) was measured by three different physicians using Orbscan II, Pentacam HR and Sirius. The normality of the distribution was evaluated by with Kolmogorov-Smirnov test. The correlations between CCT obtained from each device and refractive defect and age were evaluated using the Pearson test. The differences were evaluated by the Student paired t-test using SPSS 18.0 (IBM Corp. Armonk, New York). RESULTS: Orbscan II provided significant (p<0.0001) lower CCT measurements then both Pentacam HR (-13.66±16.53μm) and Sirius (-15.18±17.16μm); Sirius showed values slightly higher than Oculus Pentacam HR (+1.52±6.21μm) that appeared to be statistically significant (p<0.015). CONCLUSIONS: The measurement of CCT by Sirius and Pentacam HR provides similar results. By contrast, the results obtained by Orbscan II are different from those obtained from both Sirius and Pentacam HR. Copyright © 2015 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.
Purpose: To compare corneal pachymetry values measured by three different optical devices: Orbscan II, Pentacam HR and Sirius in healthy eyes. Methods: The central corneal thickness (CCT) of 102 eyes of 102 healthy subjects (mean age of 33.09. ± 8.72 years and mean refractive defect -4.11. ± 4.74 D) was measured by three different physicians using Orbscan II, Pentacam HR and Sirius. The normality of the distribution was evaluated by with Kolmogorov-Smirnov test. The correlations between CCT obtained from each device and refractive defect and age were evaluated using the Pearson test. The differences were evaluated by the Student paired t-test using SPSS 18.0 (IBM Corp. Armonk, New York). Results: Orbscan II provided significant ( p< 0.0001) lower CCT measurements then both Pentacam HR (-13.66 ± 16.53 μm) and Sirius (-15.18 ± 17.16. μm); Sirius showed values slightly higher than Oculus Pentacam HR (+1.52 ± 6.21. μm) that appeared to be statistically significant ( p< 0.015). Conclusions: The measurement of CCT by Sirius and Pentacam HR provides similar results. By contrast, the results obtained by Orbscan II are different from those obtained from both Sirius and Pentacam HR.
Central corneal thickness evaluation in healthy eyes with three different optical devices
LANZA, Michele
Conceptualization
;LANZA, Alessandro;BIFANI SCONOCCHIA, Mario
2015
Abstract
Purpose: To compare corneal pachymetry values measured by three different optical devices: Orbscan II, Pentacam HR and Sirius in healthy eyes. Methods: The central corneal thickness (CCT) of 102 eyes of 102 healthy subjects (mean age of 33.09. ± 8.72 years and mean refractive defect -4.11. ± 4.74 D) was measured by three different physicians using Orbscan II, Pentacam HR and Sirius. The normality of the distribution was evaluated by with Kolmogorov-Smirnov test. The correlations between CCT obtained from each device and refractive defect and age were evaluated using the Pearson test. The differences were evaluated by the Student paired t-test using SPSS 18.0 (IBM Corp. Armonk, New York). Results: Orbscan II provided significant ( p< 0.0001) lower CCT measurements then both Pentacam HR (-13.66 ± 16.53 μm) and Sirius (-15.18 ± 17.16. μm); Sirius showed values slightly higher than Oculus Pentacam HR (+1.52 ± 6.21. μm) that appeared to be statistically significant ( p< 0.015). Conclusions: The measurement of CCT by Sirius and Pentacam HR provides similar results. By contrast, the results obtained by Orbscan II are different from those obtained from both Sirius and Pentacam HR.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.